A 0.05°C thermal-sensitivity figure does not establish ±0.05°C temperature-measurement accuracy or suitability for medical diagnosis. This page does not supply a model-specific test report or evidence of an authorized medical intended use for a Phonemax camera.
Ask what the number measures
If 0.05°C describes a temperature difference, it equals 50mK. Confirm whether the supplier means noise equivalent temperature difference (NETD), a displayed increment or measurement accuracy. These descriptions must not be used interchangeably.
FLIR explains NETD as a signal-to-noise measure of thermal sensitivity, distinct from accuracy. A claimed sensitivity value should be accompanied by its test conditions; it does not document a clinical outcome.
Separate the evidence needed for each claim
| Claim | Evidence to request |
|---|---|
| Thermal sensitivity | A defined metric, test conditions and report for the exact imaging configuration. |
| Temperature accuracy | A separate accuracy specification and supporting measurement report. |
| Medical intended use | The manufacturer's intended-use documentation, applicable authorization and validation for that use. |
Keep clinical decisions outside this checklist
The FDA has warned that improperly used thermal imaging systems can produce inaccurate temperature readings. This article provides no diagnostic procedure. Do not use a sensitivity number or an ordinary thermal-phone image to make a medical diagnosis.
Questions before approving a purchase
- Which model, lens, firmware and application were evaluated?
- Does the report support the same metric stated in the quotation?
- Is the proposed medical use actually covered by the device documentation?
- Who in the purchasing organization will assess the technical and clinical evidence?
When evidence is missing, retain the requirement as unverified. A sharper-looking image does not fill a gap in accuracy or intended-use documentation. This is a specification-review guide, not a claim that Phonemax supplies a validated medical thermal camera.
Primary sources
FLIR: detector sensitivity and NETD; FDA: improper use of thermal imaging devices (2021 safety context).